Start-up - How to install the RF handset: The TD3 can accept commands from up to 2 RF handsets. From the factory the TD3 cannot recognise any RF handsets at all and must be matched to the RF handset(s) by the end-user/ bed manufacturer before these can control the TD3 operation. Conditions which must be fulfilled prior to activating the learning mode: • Both the back and the leg actuators must be in the fully down position; if one or both actuators are NOT in the fully down position; the matching procedure cannot be continued as now an ordinary reset down function is performed instead. Thereafter, the reset key must be released and reactivated. Activating the learning mode: • Activate the reset key on the TD3 by using a pen or similar to keep the button pressed. • Keeping the reset key activated; the RF handset(s) must be activated one at a time by pressing a random key on the RF handset. The RF handset IDs are stored in the TD3 memory and at the same time, previous RF handset IDs are erased. HB12RF HB22RF HC20RF • After having activated the RF handset keys; the reset key must be released. • If no RF handset keys are activated during the matching procedure; no changes are made in the ID memory of the TD3. Page 14 of 40