Introduction QLCI2 | 3 The Quad Load Cell Interface 2 (QLCI2) for the hospital and care segment is a scale system accessory with weighing capabilities and Out of Bed functionality. The housing makes the QLCI2 easy to mount by unique slide-on brackets and has an IPX6 Washable DURA TM ingress protection. The QLCI2 supports the LINAK OpenBus TM system, offering a high level of customisation. LINAK A/S delivers the OpenBus system only and is not responsible for any products (i.e. products from 3rd party suppliers) or the compatibility of such products with the LINAK OpenBus system. Features and options Housing colour: Protection class: Weight: Power consumption: Mounting: Cable lock: OpenBus interface: Calibration interface: Load cell interface: Optional relay gateway: Standby power: Power supply: Operation mode: Weighing precision: Light grey, RAL 7035 IPX6 Washable DURA ≈ 400 g V-Permanent 40 V = < 50 mA V-Bus 8 V = < 15 mA Mounting bracket or bed frame Integrated in lid 1 port - 10-pin modular plug 1 port - 6-pin Micro-Fit 4 ports - 6-pin Micro-Fit 6-pin Mini-Fit with two relays NC-NO There is no standby consumption if the system is on sleep Via OpenBus™ On mains or on battery 100 g accuracy