BA16 | 3 Introduction The battery box BA16 is developed for use together with the LINAK CA and CO control box series to support power backup. The BA16 has an intelligent LED indicator that indicates the battery status. The control box can monitor the voltage of battery in real-time Features and options Lead acid technology: Lead acid Output voltage: 24 V DC Max. discharge current: 6 A Charging current: 300 mA, controlled by BA16 internal charger Standby power consumption: <25 uA/<0.8 mW Capacity: 1.3 Ah/24 V Weight: 1513 g Material: PP Housing colour: Light grey (RAL 7035) Protection class: IPX6 Washable DURA™ Packaging: Multibox PCP version: PCP 2.0 Usage Compatibility: Duty cycle: Charging: Charging time: Recharging during storage: Operating temperature: Storage temperature: Relative humidity: Atmospheric pressure: Height above sea level: Service: Approvals: Battery back-up for COXX and CAXX 10%, 2 minutes continuous use followed by 18 minutes not in use Via integrated charger Approx. 6 hours Battery recharging no later than 6 months after production +5 °C to +40 °C -10 °C to +50 °C The batteries must be stored in an applicable storage room 20% to 80% - non-condensing 700 to 1060 hPa Max. 3000 meters Battery cells cannot be replaced as the battery cover cannot be closed properly afterwards. IEC60601-1 ANSI/AAMI ES60601-1 CAN/CSA-22.2 No. 60601-1 UL tested in accordance with UL60601-1