DIMENSIONS AND WEIGHTS (NOMINAL) (measured in mm unless noted) Model Description L1 L2 H1 H2 B Weight (kgs.) no. (depth) with and without tubes and needles without rubber protection boot 85 117 159 189 31 FAC6HP with rubber protection boot 98 123 173 203 42 1.25 / 0.46 1.41 / 0.62 H2 H1 MODEL NUMBER SELECTION FlowCon Commissioning Meter L1 L2 __ F _ A _ C __ 6 _ H __ P __ FUNCTIONS Battery: 9V 6LR61 Connection to PC requires optional software (see the accessories): USB Mini Flow characteristics for: FlowCon Composite Insert series (20-40mm) FlowCon E-JUST Insert series (20-50mm) FlowCon Stainless Steel Insert series (20-80mm) FlowCon SH series (15-150mm) FlowCon Green Insert series (20-40mm) FlowCon SME Insert series (20-40mm) Note: Measurement only to be done either without actuator or with actuator fully open FlowCon SM series (15-150mm) Note: Measurement only to be done either without actuator or with actuator fully open Measurement range: 0.5-600 kPaD Measurement accuracy: ± 0.2 kPaD within 0.5-20 kPaD range ± 1.0% of reading within 20-600 kPaD range Hysteresis: ± 1.0% of reading within 20-600 kPaD range Storage: Default 100 locations. Up to 2000 locations with optional software (see the accessories options) DP units: Pa, kPa, bar, m H 2 O, mm H 2 O, cm H 2 O, psi, feet H 2 O, inches H 2 O Flow units: l/s, l/min, l/h, (imperial) gpm, (US) gpm. Tech note · FlowCon FAC6HP commissioning meter · february 2016 · www.flowcon.com Page 2 of 4