LA31 | 5 Usage • Duty cycle: Max 10% or 2 minutes continuous use followed by 18 minutes not in use. • Cycles: The LA31 life cycle test has been performed with a stabilised power supply (10% duty cycle) on a 200 mm stroke actuator at max. load for 10,000 cycles (at ambient temperature) • Ambient temperature: +5° to + 40°C • Compatibility: Compatible with LINAK control boxes. Please contact LINAK. • Approvals: IEC 60601-1, ANSI/AAMI ES60601-1 and CAN/CSA-22.2 No 60601-1